Mercury specification CV-AFS Tekran 2500

Page author : Angélique Carrara.

The Tekran 2500 CV-AFS is part of the ’Mercure’ analytical park.
This park offers a wide range of analyses of the different phases of mercury (dissolved, solid and gaseous), inorganic and organic. The park consists of 3 total dissolved mercury analysers, 2 total dissolved methylmercury analysers, 2 total solid mercury analysers and 3 gaseous mercury analysers. These tools allow the speciation of mercury at ultra-trace levels (< absolute g) and in various matrices.
The CV-AFS Tekran 2500 are cold vapor atomic fluorescence spectrophotometers. They continuously measure elemental mercury under inert gas flow.

TEKRAN 2500
Framework for use Apparatus used in the framework of research projects at ISTerre and OSUG, or in the framework of external services.
Technical Specifications Sample Types
Aqueous, with preliminary sampling under ultra-clean conditions (contact us beforehand).

Technical data
Treatment of samples by bromination and tin chloride.
Method by generation of Hg0 gas (Cold Vapor), trapping and detection by Tekran 2500 Atomic Fluorescence Spectrometry (AFS).
LD : 50 pg/L.

Types of analysis
Total mercury THg, dissolved gaseous mercury DGM, ’reactive’ mercury THgR.
Quality Control US-EPA Method 1631.
Certified standards ORMS-5, NIST.
International intercalibrations GEOTRACES.
Tool Manager(s),
Team contact


 Scientific Director
Stephane Guedron 04 76 63 59 28