. 12C, 13C CM-CRDS PICARRO
Auteur de la page : Sylvain Campillo
δ13C is a major proxy used in :
- geochemistry
- paleoclimatology
- paleoceanography
It varies over time as a function of :
👉 primary productivity
👉 carbon burial
👉 vegetation type
δ13C corresponds to the ratio of stable carbon isotopes
13C / 12C, expressed in ‰ (per mil).
δ13C is expressed relative to an international standard :
👉 VPDB – Vienna Pee Dee Belemnite
It corresponds to the unusually high (13C / 12C) ratio (0.0112372) of a Cretaceous fossil belemnite shell, originating from the Pee Dee Formation (South Carolina).
CM-CRDS is dedicated to the measurement of stable carbon isotopes
12C and 13C from solid samples.
🔹 Analytical capabilities :
total carbon measurement
total carbon / organic carbon discrimination
👉 via sample decarbonation
(method developed in the laboratory)
🔹 Output provided :
%C bulk or OM and δ13C value expressed relative to the VPDB standard
🔧 CRDS system diagram
| 🧪 CM-CRDS PICARRO | |
|---|---|
| Operating framework | Instrument mainly used for research projects at ISTerre and OSUG, and more rarely for external analytical services. |
| Sample types | Solids, with specific preliminary preparation |
| Technical data | Analysis time : 10 minutes / sample |
| Capacity | 33 samples / session, 10 standards, 7 blanks |
| Concentration range | 0.5 to 1.5 mg of pure carbon / sample |
Flash combustion converts carbon into 12CO₂(g) and 13CO₂(g).
Other compounds are trapped to prevent entry into the resonant cavity.
Detection is based on measurement of laser intensity decay within the cavity
(absorbance, ppmv).
| Types of analyses | |
|---|---|
| Carbon 12C and 13C | |
| Isotopic analyses on decarbonated samples (method developed in the laboratory to extract organic matter) |
✔ High-precision measurement of stable carbon isotopes
✔ Robust and reproducible method
✔ Ability to distinguish total carbon from organic carbon
✔ Locally developed methodology
✔ Full integration within the ISTerre analytical platform
👉 For any analysis request, please complete and return the form below, signed, to Sylvain Campillo.
⚠️ Analyses may be refused if :
sample quality
or preparation conditions
are not considered adequate.
📌 To avoid any issues, it is strongly recommended to contact the responsible scientists during the project design phase.
| Quality control | Certified standards |
|---|---|
| Prior contact mandatory |
| CRDS contacts | |
|---|---|
| Stéphane Guédron 📞 +33 4 76 63 59 28 | |
| Sylvain Campillo |